电感耦合等离子体发射光谱法测定工业废水中32种元素

  来源:《现代科学仪器》期刊2017-01-23点击:408


【核心介绍】本文根据HJ 776-2015测定工业废水中32种元素总量,测试结果表明:ICP-OES法检出限低,轴向测量、侧向测量、轴向测量plus、侧向测量plus,4种测量方式满足4个数量级浓度线性范围要求;线性拟合系数均在 R=0.9990~1.0000;精密度RSD均≤1%;加标回收率以及质控样回收率均在90%-110%,测试结果满意。ICP-OES光学分辨率高,可选用的无干扰分析线多,实现多元素同时无干扰的测量。  

电感耦合等离子体发射光谱法测定工业废水中32种元素

沙德仁 杨静*

(德国耶拿分析仪器股份公司)

摘要本文根据HJ 776-2015测定工业废水中32种元素总量,测试结果表明:ICP-OES法检出限低,轴向测量、侧向测量、轴向测量plus、侧向测量plus,4种测量方式满足4个数量级浓度线性范围要求;线性拟合系数均在 R=0.9990~1.0000;精密度RSD均≤1%;加标回收率以及质控样回收率均在90%-110%,测试结果满意。ICP-OES光学分辨率高,可选用的无干扰分析线多,实现多元素同时无干扰的测量。

关键词高分辨电感耦合等离子体发射光谱;工业废水;测定多种化学元素

中图分类号 X703


Determination of 32 Elements in The Industrial Wastewater by ICP-OES

Sha Deren  Yang Jing*

(Analytik Jena AG)

Abstract According to the HJ 776-2015,We determined the total of 32 elements in the industrial wastewater. The results showedthat we get the lower detection limit, and the requirements of 4 orders of magnitude concentration of linear range were met by the axial, lateral, axial plus, the lateral plus 4 kinds of measurement methods; the linear correlation R isfrom 0.9990 to 1.0000,RSD≤1%; the recovery of addition and QC are from 90% to 110%, the results is satisfied. With the high optical resolution,  multi-lines can be used , and fast sequential and simultaneous multi-element analysis can be realized without interference.

Key words High Resolution ICP-OES; Industrial wastewater; Determination of various chemical elements


水污染是我国面临的主要环境问题之一,随着我国工业的发展,工业废水的排放量日益增加,严重污染地表水和地下水,影响工农渔业生产,直接或间接地给人民生活和身体健康带来危害。而重金属(如含镉、镍、汞、锌等)废水是对环境污染最严重和对人类危害最大的工业废水之一,如何有效地评价和治理重金属污染已成为人类共同关注的问题。本文按照《水质32种元素的测定电感耦合等离子体发射光谱法》(HJ776-2005)[1],采用电感耦合等离子体发射光谱(ICP-OES)法测定工业废水中的有害元素,检出限低,稳定性好,线性范围宽,测试结果准确,同时测量元素种类多,检测速度快,效果满意。

1 实验条件

1.1 仪器设备

PQ9000高分辨率电感耦合等离子体发射光谱仪(德国耶拿分析仪器股份公司):采用蔡司技术光学系统,棱镜、中阶梯光栅两级色散,光学分辨率0.003nm,波长范围160900nm,波长准确度0.0004nm。垂直炬管,双向观测,轴向、轴向扩展、侧向、侧向扩展4种测量方式,适合各类样品(有机、高盐)的分析,满足不同浓度的同时测量。高量子效率和紫外高灵敏度的新一代CCD检测器,像素分辨率0.002nm,同时记录元素线与其直接光谱环境,自动扣除背景。自激式、40.68MHz、0.7~1.7kW功率可调射频发生器,各路气体均用质量流量控制器(MFC)控制,等离子体持续稳定。吹扫用氩气又回到等离子体再用,既能持续吹扫,提高灵敏度,又不额外消耗氩气,运行成体低。组合式炬管,卡插式定位,操作方1.2 仪器条件

等离子体强度:1200W;等离子气流量:12.0L/min;辅助气流量:0.5 L/min;雾化气流量:0.5 L/min。

1.3 试剂和材料

硝酸:高纯,高氯酸:优级纯,水:一级,氩气:高纯。

Li、Na、K、Be、Mg、Ca、Sr、Ba、B、Al、Si、Sn、Pb、P、As、Sb、Bi、S、Se、Ti、Zr、V、Cr、Mo、Mn、Fe、Co、Ni、Cu、Ag、Zn、Cd单一标准储备溶液:1000mg/L。

XCCC-13A-500: Li、Na、K、Be、Mg、Ca、Sr、Ba、B、Al、Pb、P、As、Bi、Se、V、Cr、Mn、Fe、Co、Ni、Cu、Zn、Cd,24种元素混合标准储备溶液:1000 mg/L。

XCCC-14B-500: Si、Sn、Sb、S、Ti、Mo、W、Re、Pd 9种元素混合标准储备溶液:1000 mg/L。

1  32种元素标准溶液浓度范围

元  素

标准及C(mg/L)

标准要求

0

1

2

3

4

5

Ag

0.0~500.0

0.0

1.0

10.0

100.0

300.0

500.0

Li Na K Ca Ba Al

0.0~250.0

0.0

0.5

5.0

50.0

150.0

250.0

Be Mg Sr B

0.0~500.0

0.0

1.0

10.0

100.0

300.0

500.0

Fe

0.0~250.0

0.0

0.5

5.0

50.0

150.0

250.0

Pb Cr Mn Co Ni Cu Zn Cd

0.0~500.0

0.0

1.0

10.0

100.0

300.0

500.0

Ti

0.0~250.0

0.0

0.5

5.0

50.0

150.0

250.0

Sn As Sb Bi Se V Mo Zr

0.0~500.0

0.0

1.0

10.0

100.0

300.0

500.0

Si

0.0~250.0

0.0

0.5

5.0

50.0

150.0

250.0

P S

0.0~500.0

0.0

1.0

10.0

100.0

300.0

500.0


1.4干扰试验

P标准曲线:0.0、1.0、10.0、100.0mg/L1%(v/v)硝酸。在4份10 mg/L P标准溶液中分别加入1.0、10.0、100.0、500.0 mg/L Fe标准溶液,1%(v/v)硝酸。

Zn标准曲线:0、1.0、10.0、100.0mg/L1%(v/v)硝酸。在4份10 mg/LZn

标准溶液中分别加入1.0、10.0、100.0、500.0 mg/L Cu标准溶液,1%(v/v)

硝酸。

2  样品处理

2.1样品采集

按照HJ/T91[2]HJ/T164[3]的相关规定进行水样的采集。采样前,用洗涤剂和一级水依次洗净聚乙烯瓶,置于硝酸溶液浸泡24h以上,用实验用水彻底洗净。样品采集后立即加入适量硝酸,使硝酸含量达到1%。

2.2样品制备

50.0mL某化工厂工业废水样品加入5.0mL硝酸,置于电热板上加热消解,在不沸腾的情况下,缓慢加热至近干。取下冷却,反复进行这一过程,直至试样溶液颜色变浅或者稳定不变。冷却后,加入硝酸若干毫升,再加入少量水,置电热板上继续加热使残渣溶解。冷却后,用实验用水定容至原取样体积,使溶液保持1% 的硝酸酸度。

注:对于某些基体复杂的废水,消解时可加入2.0-5.0mL高氯酸消解。若消解液中存在一些不溶物,可静置或离心分离10min以获得澄清液。(若离心或静置过夜后仍有悬浮物,则可过滤去除,但应避免过滤过程中可能的污染。)

2.3空白样品

用一级水代替样品,按与样品制备相同的步骤进行空白试样的制备。

3  样品测定

3.1 32元素标准曲线线性方程、相关系数、精密度

32元素绘制标准曲线,标准曲线线性相关系数在0.9990~1.0000,精密度≤1.00%。( 重复3次测量)


232元素标准曲线线性方程、相关系数、精密度

元素

谱线nm

线性方程

相关系数(R2

RSD %(标准2)

Be

249.458

y=7065.06+13230.76x

0.9999

0.49

Sr

216.596

y=16949.46+10228.57x

0.9997

0.42

Ba

233.527

y=10083.60+17964.62x

0.9998

0.41

Mg

279.078

y=2065.87+6684.45x

0.9999

0.33

Ca

315.887

y=12815.40+27276.01x

0.9999

0.75

B

249.678

y=-12487.19+19060.02x

0.9999

0.99

Al

396.152

y=10755.96+28089.41x

0.9999

0.80

Na

589.592

y=7658.32+9495.63x

0.9999

0.37

Li

670.791

y=70704.84+48843.63x

0.9991

0.70

K

769.897

y=4280.31+9326.18x

0.9999

0.46

Ag

328.068

y=15597.57+12800.34x

0.9998

0.37

Cd

228.802

y=13857.18+3419.88x

0.9991

0.21

Fe

259.940

y=4189.20+16301.48x

0.9999

0.49

Co

240.725

y=791.35+4913.38x

0.9999

0.44

Cr

425.433

y=5707.77+10753.96x

0.9999

0.71

Cu

327.396

y=856.44+13495.58x

0.9999

0.36

Mn

293.306

y=1738.25+5328.15x

1.0000

0.66

Ni

351.505

y=-1426.42+5828.15x

1.0000

0.31

Pb

220.353

y=-1052.05+3368.90x

1.0000

0.59

Zn

206.200

y=5007.88+5714.84x

0.9998

0.90

As

188.979

y=16348.62+5643.16x

0.9991

0.33

Sb

217.581

y=6995.72+4180.71x

0.9993

0.74

Bi

223.061

y=7497.67+5065.92x

0.9993

0.68

Se

196.028

y=10775.24+4606.98x

0.9991

0.94

Mo

202.030

y=-1632.18+2615.74x

0.9999

0.50

V

370.357

y=-8479.27+6040.25x

0.9998

0.32

Sn

286.332

y=-2147.33+3032.62x

0.9999

0.78

Ti

308.804

y=8779.98+22950.30x

1.0000

0.51

Zr

273.486

y=-687.51+4448.49x

0.9999

0.47

S

182.565

y=-1488.00+612.22x

0.9992

0.42

P

213.618

y=-1038.05+2909.04x

1.0000

0.94

Si

251.611

y=21559.23+17940.89x

0.9989

0.34

3.2 工业废水样品测定结果

对空白溶液同时测定11次,取3倍标准偏差所对应的浓度为仪器的检出限,结果如下表所示,远低于HJ 776-2015 《水质 32种元素的测定 电感耦合等离子体发射光谱法》中对检出限的要求,对工业废水样品进行加标回收试验,回收率在90%-110%,对质控样品XCCC进行测定,回收率在90%-110%,结果满意,结果如表3所示。


3工业废水样品测定检出限,样品含量,加标回收率以及质控样品回收率

元素

样品

检出限

mg/L

含量

mg/L

加标量

mg/L

加标

测量值

mg/L

回收率(RSD:%)

质控样品

含量(mg/L

标准定值

mg/L

回收率(RSD:%)

Be

工业废水

样品

0.00004

0.02211

1.0

1.02110

94.9%

X

C

C

C

49.448

50 mg/L

98.9%

Sr

0.0006

0.0661

1.0

1.0642

96.6%

49.810

99.6%

Ba

0.0002

0.0232

0.5

0.5212

93.4%

49.196

98.4%

Mg

0.0002

0.9794

1.0

2.0371

105.9%

50.452

100.9%

Ca

0.0006

5.6683

0.5

6.0772

98.4%

51.176

102.4%

B

0.0006

0.3193

1.0

1.2964

92.9%

51.441

102.9%

Al

0.0005

0.4192

0.5

0.8961

94.4%

49.223

98.4%

Na

0.01

0.67

0.5

1.16

97.9%

51.461

102.9%

Li

0.003

0.018

0.5

0.520

109.3%

54.641

109.3%

K

0.0003

0.6091

0.5

1.1494

106.5%

52.734

105.5%

Ag

0.0004

0.0082

0.5

0.5075

92.3%

---

---

Cd

0.0005

0.0111

1.0

1.0121

108.4%

51.689

103.4%

Fe

0.003

0.105

0.5

0.599

94.4%

49.192

98.4%

Co

0.001

0.048

1.0

1.044

91.4%

49.203

98.4%

Cr

0.003

0.098

1.0

1.106

108.6%

51.781

103.6%

Cu

0.0006

0.0886

1.0

1.0844

95.6%

49.308

98.6%

Mn

0.0006

0.0785

1.0

1.0804

103.0%

50.509

101.0%

Ni

0.0003

0.0889

1.0

1.0971

108.8%

51.908

103.8%

Pb

0.0017

0.0254

1.0

1.0234

92.6%

46.282

92.6%

Zn

0.0001

0.0333

1.0

1.0310

95.3%

47.673

95.3%

As

0.0030

0.0664

1.0

1.0713

107.7%

50.855

101.7%

Sb

0.0035

0.2818

1.0

1.2988

105.9%

52.450

104.9%

Bi

0.0026

0.0424

1.0

1.0405

94.7%

47.370

94.7%

Se

0.0044

0.1111

1.0

1.1064

95.5%

49.235

98.5%

Mo

0.001

0.056

1.0

1.061

108.1%

50.572

101.1%

V

0.0004

0.0389

1.0

1.0428

109.8%

50.879

101.8%

Sn

0.0053

0.0909

1.0

1.0864

94.0%

51.979

104.0%

Ti

0.0001

0.1017

0.5

0.5948

93.2%

51.586

103.2%

Zr

0.00036

0.25904

1.0

1.27316

105.4%

---

---

S

0.018

0.190

1.0

1.202

106.4%

51.678

103.4%

P

0.0073

0.3562

1.0

1.3374

94.6%

47.294

94.6%

Si

0.0013

1.8209

0.5

2.2094

93.9%

49.451

98.9%


4  Fe对P 213.6182nm的干扰验证

4  Fe对P 213.6182nm的干扰验证实验

P 213.6182nm

Fe

测量值

10.0ppm

0.0 ppm

10.9009 ppm

10.0ppm

1.0 ppm

10.8652 ppm

10.0ppm

10.0 ppm

10.9710 ppm

10.0ppm

100.0 ppm

10.5902 ppm

10.0ppm

500.0 ppm

10.6508 ppm



1  100ppmFe对10ppmP的干扰验证谱图



计算公式: K=


K-干扰系数

cm–净干扰浓度

cn–净分析浓度

M–被干扰元素浓度

N–干扰元素浓度

计算结果:以直接比较法计算干扰系数K1=0.230,K10=0.0230,K100=0.00300,

K500=0.000120。

无论从数据还是峰图看,Fe均未对P 213.6182nm构成干扰,只是当Fe的浓度到100 ppm时,P左右的两个Fe峰初露,Fe的浓度达到500ppm时,P左右的两个Fe峰显现了,但与P 的213.6182nm完全分开,不存在干扰。

5  结论

1)从测试数据看,标准曲线的浓度低,线性范围宽,达到4个数量级,可满足各种工业废水样品含量浓度范围宽的测试;RSD均≤1.0%;线性拟合好:R=0.9990~1.0000;可选谱线多,充分表征了ICP-OES法的高分辨率,高灵敏度和高稳定性。

2)从测试结果看,本法加标回收率以及质控样品的回收率均在90.0%-110.0%,结果准确。

3)干扰试验实验证明,Fe未对P的213.6182nm产生干扰,高分辨率ICP-OES法能对每个元素提供多条没有光谱干扰的灵敏度不同的分析线,尤其有利于工业废水等背景复杂样品的分析,无需做校正系数,使检测准确可靠,简便快速。